KALTEC SCIENTIFIC, INC. |
||||||||||
|
||||||||||
![]() |
Bob does not rotate when running a test. Likely Cause: DC fuse (located in back of unit) has blown. Solution: Replace with AGC 20 fuse only. Power switch indicator or LEDs do not light up when the power is turned on (and unit is properly plugged in). Likely Cause: AC fuse is blown. Solution: Contact Kaltec's Service Department before replacing. >> Return to top << Drive Enabled LED does not illuminate. Torque Overload LED is on and sounder is making noise. Likely Cause: Emergency Stop button has been pushed. Solution: Deactivate Emergency Stop button by pushing on it and rotating it counterclockwise. Likely Cause: Full scale torque has been reached. Solution: Switch viscometer off for at least two seconds, then on again. >> Return to top << Viscosity and/or Torque values are off 30-5%. In WinShear under the SETTINGS menu, select MACHINE. If your viscometer is 6600 RPM, enter the spindle ratio of 2. If your viscometer is 8800 RPM, enter the spindle ratio of 3. |
|||||||||
Kaltec Scientific, Inc. |
© 2009 Kaltec Scientific, Inc. |
|||||||||